Most powerful Multimode inlet for GC and GCMS systems
The 4th generation of the OPTIC multimode injection system is a versatile
tool for the Shimadzu GC and GCMS series which increases application range and detection
limits.
Wider application range for gas chromatography
In addition to the standard sampling modes the programmable OPTIC-4 can be operated
as Large Volume injector, Pyrolysis or Thermal Desorption introduction system. DMI
(Difficult Matrix Introduction) is another option where µ-vials filled with sample
(solid or highly matrix loaded samples) can be introduced into the OPTIC-4 liner
and allow direct analysis of the compounds out of the sample.
The patented low thermal mass design together with the direct resistive heating
provides linear temperature range up to 600°C at rates as high as 60 °C/sec and opens
the possibility to apply the OPTIC-4 for this wide range of applications. In combination
with the auto sampler AOC-5000 Plus the sample introduction can be fully automated
with an automatic liner exchanger (LINEX). In addition the variety of different
liners supports the huge range of applications using the OPTIC-4.
Improved detection limits
By using the OPTIC-4 Injector as cold injection system also compounds that normally
degrade in hot injectors e.g. thermal labile pesticides can be analyzed without
difficulties. Even for low boiling point compounds precise and reproducible results
can be obtained. The temperature range of the OPTIC-4 is +10 °C above RT to 600 °C
(-180 °C or -50 °C to 600 °C by adding the injector cooling option with direct LN2
or CO2 cooling, respectively).
Large Volume injections improve the detection limits and can be used in all application
areas where sensitivity is crucial e.g. in environmental or food analysis. The unique
solvent monitor enables easy method development for Large Volume injections.