| Wavelength range |
190 to 1100nm |
| Spectral bandwidth |
1nm (190 to 1100nm) |
| Wavelength display |
0.1-nm increments |
| Wavelength setting |
0.1-nm increments (1-nm increments when setting scanning range ) |
| Wavelength accuracy |
±0.1nm @ 656.1nm D2 ±0.3nm (190 to 1100nm) |
| Wavelength repeatability |
±0.1nm |
| Stray light |
less than 0.02% NaI @ 220nm, NaNO2 @ 340nm |
|
less than 1.0% KCl @ 198nm
|
| Photometric system |
Double Beam |
| Photometric range |
Absorbance: -4to4 Abs Transmittance: 0% to 400% |
| Photometric accuracy |
±0.002 Abs (0.5Abs) |
|
±0.004 Abs (1.0Abs) |
|
±0.006 Abs (2.0Abs) |
| Photometric repeatability |
less than ±0.001 Abs (0.5Abs) |
|
less than ±0.001 Abs (1Abs) |
|
less than ±0.003 Abs (2.0Abs) |
| Baseline stability |
less than 0.0003 Abs/H @700nm (one hour after light source turned ON) |
| Baseline flatness |
within ±0.0006 Abs (190 to 1100nm,one hour after light source turned ON) |
| Noise level |
within 0.00005 Abs RMS value (@700nm) |
| Dimensions (W×D×H) |
450(W) x 490(D) x 270(H) |
| Weight |
15kg
|
| Printers |
DPU, ESC/P, PCL printers, USB I/F Windows-compliant printers
are available with USB memory and PC software
|
| Memory |
USB memory (option) Saved as text and UVPC file |
| Performance for PC
|
USB memory + UVProbe (standard) Win XP
|